Specifications and Reference Data
March 2016
Reference Manual
00809-0100-4772, Rev FA
110
Specifications and Reference Data
Process temperature limits
Standard
–40 to 450 °F (–40 to 232 °C)
Extended
–330 to 800 °F (–200 to 427 °C)
Severe
–330 to 842 °F (–200 to 450 °C)
-157 to 800 °F (-105 to 427 °C) for European Pressure
Equipment Directive (PED), consult factory for lower
operating temperatures.
The flowtube and sensor, in remote mount
configurations, is functionally rated to 842 °F (450 °C)
process temperature. Process temperature may be
further restricted depending on hazardous area options
and PED certificates. Consult applicable certificates for
particular installation limits.
Multivariable (MTA option)
–50 to 800 °F (–45 to 427 °C)
Use above 450 °F (232 °C) requires Extended Sensor
Output signal
Completely digital output with F
OUNDATION
Fieldbus
communication (ITK 6.0 compliant).
FOUNDATION
Fieldbus transmitter power
requirement
Power consumption
Based on maximum ratings=600 mWatts
Ambient temperature limits
Operating
–58 to 185 °F (–50 to 85 °C)
–4 to 185 °F (–20 to 85 °C) for flowmeters with local
indicator
Storage
–58 to 250 °F (–50 to 121 °C)
–50 to 185 °F (–46 to 85 °C) for flowmeters with local
indicator
Pressure limits
Flange style meter
Rated for ASME B16.5 (ANSI) Class 150, 300, 600, 900,
and 1500, EN 1092-1, PN 10, 16, 25, 40, 63, 100, and
160, and JIS 10K, 20K, and 40K
Reducer style meter
Rated for ASME B16.5 (ANSI) Class 150, 300, 600, and
900, EN 1092-1, PN 10, 16, 25, 40, 63, 100, and 160.
Dual sensor style meter
Rated for ASME B16.5 (ANSI) Class 150, 300, 600, 900,
and 1500, EN 1092-1, PN 10, 16, 25, 40, 63, 100, and
160, and JIS 10K, 20K, and 40K
Wafer style meter
Rated for ASME B16.5 (ANSI) Class 150, 300, and 600,
EN 1092-1, PN 10, 16, 25, 40, 63, and 100, and JIS 10K,
20K, and 40K
Note
All wafer style meters are pressure rated and labeled
at 1500 PSI / 10.34 MPa at 100 °F / 38 °C regardless of
alignment ring size code ordered.
Weld-end style meter
W1 Welds to Schedule 10 mating pipe
Max Working Pressure 720 psig (4.96 MPa-g)
W4 Welds to Schedule 40 mating pipe
Max Working Pressure 1440 psig (9.93 MPa-g)
W8 Welds to Schedule 80 mating pipe
Max Working Pressure 2160 psig (14.9 MPa-g)
W9 Welds to Schedule 160 mating pipe
Max Working Pressure 3600 psig (24.8 Mpa-g)
Table A-2. Minimum Measurable Meter
Velocities
(1)
Feet per Second
Meters per
Second
Liquids
(2)
Gases
(2)
The ρ is the process fluid density at flowing
conditions in lb/ft
3
for ft/s and kg/m
3
for m/s
1. Velocities are referenced to schedule 40 pipe.
2. This minimum measurable meter velocity is based on default
filter settings.
Table A-3. Maximum Measurable Meter
Velocities
(1)
(Use the smaller of the two values)
Feet per Second Meters per Second
Liquids
Gases
(2)
The ρ is the process fluid density at flowing
conditions in lb/ft
3
for ft/s and kg/m
3
for m/s
1. Velocities are referenced to schedule 40 pipe.
2. Accuracy limitations for gas and steam for Dual-style meters
(
1
/2-in. to 4-in.): max velocity of 100 ft/s (30.5 m/s).
36/ρ
54/ρ
36/ρ
54/ρ
90,000/ρ or 30
134,000/ρ or 9.1
90,000/ρ or 300
134,000/ρ or 91.1
9–30 Vdc/18 mA I.S. INSTALLATIONS
9–17.5 Vdc/18 mA FISCO
9–32 Vdc/18 mA ALL OTHERS