Electronics Verification
March 2016
Reference Manual
00809-0100-4772, Rev FA
154
Electronics Verification
Figure C-12. Using Sensor Header with External Frequency Source
7. Set the sinewave wave amplitude to 2Vpp. Note: The amplitude may need to be
higher for larger line sizes and/or high density liquids.
8. Select the desired frequency(s).
9. Monitor the Flow reading on the Fieldbus host system or Field Communicator.
10. To end the external flow simulation, in the Transducer Block set the Simulation
Control to "Sim Disabled".
11. If applicable, reinstall the LCD display and fully tighten all screws.
12. Install and tighten the electronics compartment cover.