CAUTION!
EQUIPMENT DAMAGE
Do not apply 230 Vac power to a 115 Vac TCL (Model option -11). Doing so will damage
the instrument.
Power wiringFigure 4-1:
Important
Leave the pump power switches off until ready to start up the unit.
See Chapter 5.
4.3 Make power, alarm, output, and sensor
connections in the transmitter
WARNING!
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70)
and/or any other applicable national or local codes.
4.3.1 Power
Wire AC mains power supply to the power supply board, which is mounted vertically on
the left hand side of the transmitter enclosure.
The power connector is at the top of the board.
Wiring
20 Rosemount TCL