1. Resetting to factory default clears ALL user-entered settings, including sensor and
analog output calibration, and returns ALL settings and calibration values to the
factory defaults.
2. Resetting a sensor calibration to the default value clears user-entered calibration
data for the selected sensor but leaves all other user-entered data unaffected.
3. Resetting the analog output calibration clears only the user-entered analog output
calibration. It leaves all other user-entered settings unchanged.
7.8.2 Procedure - resetting the transmitter
Complete the following steps to reset the transmitter.
1. Press MENU.
The main Menu screen appears.
2. Move to Program and press ENTER.
3. Scroll to the bottom of the screen and continue scrolling until Reset Analyzer is
highlighted. Press ENTER.
4. Choose whether to reset all user-entered values (Factory Defaults), sensor calibration
(Sensor Cal Only), or output calibration (Output Cal Only).
If you choose Sensor Cal Only or Output Cal Only, a second screen appears in which you
can select which sensor or output calibration to reset.
5.
To return to the main display, press MENU and then EXIT.
Programming the transmitter
Instruction Manual 57