EasyManua.ls Logo

Emerson Rosemount TCL - Page 55

Emerson Rosemount TCL
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
High Alarm LogicFigure 7-1:
The alarm activates when the chlorine concentration exceeds the high setpoint. The
alarm remains activated until the reading drops below the value determined by the
deadband.
Low Alarm LogicFigure 7-2:
The alarm activates when the chlorine concentration drops below the low setpoint.
The alarm remains activated until the reading increases above the value
determined by the deadband.
Interval timer
Any alarm relay can be used as an interval timer. Figure 7-3 shows how
the timer operates. While the interval timer is operating, the main
display, analog outputs, and assigned alarms for the sensor can be put
on hold. During hold, the main display remains at the last value.
Programming the transmitter
Instruction Manual 47

Table of Contents

Other manuals for Emerson Rosemount TCL

Related product manuals