6. Change the value in the display to match the reading from the milliameter.
If the calibration was successful, the screen below appears.
If the user entered value is more than ±1 mA different from the nominal value, a
possible error screen appears.
7.
To force the transmitter to accept the calibration, choose Yes.
8. To return to the main display, press MENU and then EXIT.
Calibration
70 Rosemount TCL