PlasmaQuant MS product family Safety instructions
19
2.7 Safety instructions - maintenance and repair
Observe the following notes:
The PlasmaQuant MS is always serviced by the service engineers of Analytik Jena or
an authorized and trained representative. Otherwise the warranty is void. The
operator may only carry out the tasks listed in the chapter "Maintenance and care".
The exterior of the PlasmaQuant MS may only be cleaned with a damp cloth. Use
only water and, if required, commercial surfactant. Do not use organic solvents or
abrasive cleaners.
Carefully close the interface door and the plasma compartment lid after
maintenance. There is a risk of your hand being crushed.
If water or other liquids are found to leak out of the instrument, contact the service
engineers.
Risk of damage to health due to improper decontamination! Perform a professional
and documented decontamination of the device before returning it to Analytik
Jena. The decontamination protocol is provided by the service when the return is
registered. Analytik Jena must refuse acceptance of contaminated devices. The
sender may be liable for any damage caused by inadequate decontamination of the
device.
Use only original spare parts, wear parts and consumables. They have been tested
and ensure safe operation. Glass parts are wear parts and are not subject to the
warranty.