10 Maintenance and care
This chapter contains all the instructions for care and maintenance which users can
and must perform themselves. All maintenance work and repairs beyond this scope
must only be performed by service engineers from Analytik Jena or persons authorized
by Analytik Jena. To maintain optimum operational performance, the
PlasmaQuant MS should be serviced on an annual basis by service engineers from
Analytik Jena. Only use replacement parts from Analytik Jena. Laboratory parts
required for routine operation can be ordered from Analytik Jena.
Analytik Jena recommends that records of the system operating conditions be
maintained. Careful comparison of records can provide valuable clues to identifying
and solving problems.
WARNING
Plasma emits UV radiation and high
-frequency electromagnetic radiation which can
cause serious eye and skin injuries as well as other health problems.
Therefore, the plasma compartment door must be closed before igniting and operating
the plasma. Safety interlocks must not be bypassed!
Before opening the plasma compartment lid or the interface door for maintenance
work, always extinguish the plasma from the software.
CAUTION
The torch compartment and interface, including the cones, are still very hot after the
plasma has been extinguished. Contact with these hot surfaces can cause burns.
Therefore, wait 5 minutes after extinguishing the plasma before touching the torch or
wea
r heat-resistant gloves.
CAUTION
Observe the safety instructions! Any unauthorized interference limits warranty
entitlements.
Compliance with these safety instructions is a requirement for error free installation
and the proper functioning of your
PlasmaQuant MS measuring environment. Always
observe all warnings and instructions which are displayed on the device itself or which
are displayed by the control program of the PlasmaQuant