Cerabar M
60 Endress+Hauser
Filling oil
When selecting the filling oil, the fluid temperature and ambient temperature as well as the operating pressure
are of crucial importance. Observe the temperatures and pressures during commissioning and cleaning. A
further selection criterion is the compatibility of the filling oil with the requirements of the fluid. For example,
only filling oils that are harmless to health - such as vegetable oil - can be used in the food industry.
→ See also the following section "Diaphragm seal filling oils".
The filling oil used influences the T
K
zero point, the temperature operating range of a diaphragm seal system
and the response time. → See also Page 61 ff, "Influence of the temperature on the zero point" section.
Pressure transmitter
The pressure transmitter influences the temperature operating range, the T
K
zero point and the response time
as a result of its volume change. The volume change is the volume that has to be shifted to pass through the
complete measuring range.
Pressure transmitters from Endress+Hauser are optimized with regard to minimum volume change.
Diaphragm seal filling oils
Version
1
Filling oil Permitted
temperature range
2
at 0.05 bar ≤ p
abs
≤ 1 bar
Permitted
temperature range
2
at p
abs
≥ 1 bar
Density Viscosity Expansion
coefficient
T
K
correction
factor
Notes
[g/cm
3
][cSt at
25°C/
77°F]
[1/K]
A, J, O, T Silicone oil –40 to +180°C
(–40 to +356°F)
–40 to +250°C
(–40 to +482°F)
0.96 100 0.00096 1 Suitable for foods
FDA 21 CFR 175.105
G, H, K, R High-
temperature oil
–10 to +200°C
(+14 to +392°F)
–10 to +400°C
(+14 to +752°F)
1.07 37 0.0007 0.72 High temperatures
N, B, C Inert oil –40 to +80°C
(–40 to +176°F)
–40 to +175°C
(-40 to +347°F)
1.87 27 0.000876 0.91 Oil for ultrapure gas
and oxygen
applications
D, F, L
3
, P, S Vegetable oil –10 to +120°C
(+14 to +248°F)
–10 to +200°C
(+14 to +392°F)
0.94 9.5 0.00101 1.05 Suitable for foods
FDA 21 CFR 172.856
E Glycerine — +15 to +200°C
(+59 to +392°F)
1.26 1118 0.000615 0.64 Suitable for foods
M, U Low
temperature oil
–70 to +80°C
–94 to +176°F
–70 to +180 °C 0.92 4.4 0.00108 1.12 Low temperatures
1) Version for feature 80 in the order code
2) Observe temperature limits of the device (→ Page 28 and Page 29)
3) Version "L" only for PMP46