Installation Cerabar S PMP71 with 4 to 20 mA HART
12 Endress+Hauser
Mount Cerabar S with shutoff device above the tapping point so that
any condensate can flow into the process.
Pressure measurement in steam
P01-PMx7xxxx-11-xx-xx-xx-002
Fig. 7: Measuring arrangement for pressure measurement in steam
1 Cerabar S
2 Shutoff device
3 U-shaped siphon
4 Circular siphon
• Mount Cerabar S with siphon above the tapping point.
The siphon reduces the temperature to almost ambient temperature.
• Fill the siphon with fluid before commissioning.
Pressure measurement in liquids
P01-PMx7xxxx-11-xx-xx-xx-003
Fig. 8: Measuring arrangement for pressure measurement in liquids
1 Cerabar S
2 Shutoff device
Mount Cerabar S with shutoff device below or at the same level as the tapping point.