Description of Device Parameters Deltabar FMD71, FMD72
100 Endress+Hauser
Min. meas.temp. (270)
Navigation Diagnosis → Sensor LP → Min. meas.temp. (270)
Write permission No write permissions. Parameter is read only.
Description Displays the smallest temperature measured in the sensor module. You can reset this value
using the "Reset peakhold (161)" parameter.
Max. meas. temp. (271)
Navigation Diagnosis → Sensor LP → Max. meas. temp. (271)
Write permission No write permissions. Parameter is read only.
Description Displays the largest temperature measured in the sensor module. You can reset this value
using the "Reset peakhold (161)" parameter.
14.11 Diagnostic → Diagnostic list
Diagnostic list
Diagnostic 1 (075)
Diagnostic 2 (076)
Diagnostic 3 (077)
Diagnostic 4 (078)
Diagnostic 5 (079)
Diagnostic 6 (080)
Diagnostic 7 (081)
Diagnostic 8 (082)
Diagnostic 9 (083)
Diagnostic 10 (084)
Navigation Diagnosis → Diagnostic list
Write permission No write permissions. Parameter is read only.
Description This parameter contains up to ten diagnosis messages that are currently pending,
arranged in order of priority.
14.12 Diagnosis → Event logbook
Event logbook