Loading...
Deltabar FMD71, FMD72 Installation
Endress+Hauser 19
2 mm
3 mm
A0017506
5.5 Thermal insulation - FMD71 high-temperature
version
The FMD71 high-temperature version may only be insulated up to a certain height. The
maximum permitted insulation height is indicated on the devices and applies to an
insulation material with a heat conductivity ≤ 0.04 W/(m x K) and to the maximum
permitted ambient and process temperature. The insulation height is not indicated on
hygienic connections.
Ambient temperature (T
A
): ≤ 70 °C (158 °F)
Process temperature (T
P
): ≤ 150 °C (302 °F)
The data were determined under the most critical application "quiescent air".
A
B
1 2
A0021075
A Ambient temperature
B Process temperature
1 Insulation height
2 Insulation material