Deltabar FMD71, FMD72 Installation
Endress+Hauser 21
L
WARNING
Material fatigue from pressure and temperature!
Risk of injury due to bursting of parts! The thread can become lose if exposed to high
pressure and temperature loads.
‣
The integrity of the thread must be checked regularly and the thread may need to be
retightened with the maximum tightening torque of 7 Nm (5.16 lbf ft). Teflon tape is
recommended for sealing the ½" NPT thread.
The mounting bracket can be installed on pipes with a diameter of 1¼" to 2" or on walls.
In the case of pipe mounting, the nuts on the bracket must be tightened uniformly with a
torque of at least 5 Nm (3.69 lbf ft).
A0017514
• The mounting bracket is included in the delivery.
• Ordering information:
Product Configurator order code for "Enclosed accessories", option "PA" or
as a separate accessory (part no.: 71102216).
5.8 Installing the transmitter
The transmitter is installed with the mounting bracket supplied. The mounting bracket can
be installed on pipes with a diameter of 1¼" to 2" or on walls.
In the case of pipe mounting, the nuts on the bracket must be tightened uniformly with a
torque of at least 5 Nm (3.69 lbf ft).
A0021145
The mounting bracket is included in the delivery.