Description of Device Parameters Deltabar FMD71, FMD72
110 Endress+Hauser
Navigation Diagnosis → Simulation → Sim. tank cont. (116)
Write permission Operators/Service engineers/Expert
Description Enter the simulation value. See also "Simulation mode".
Prerequisite "Measuring Mode" = level, Lin mode "Activate table" and "Simulation Mode" = Tank content
Value at switch-on Current tank content
Sim. current (117)
Navigation Diagnosis → Simulation → Sim. current (117)
Write permission Operators/Service engineers/Expert
Description Enter the simulation value. See also "Simulation mode".
Prerequisite "Simulation mode"= Current value
Value at switch-on Current value of the current
Sim. error no. (118)
Navigation Diagnosis → Simulation → Sim. error no. (118)
Write permission Operators/Service engineers/Expert
Description Enter the diagnostic message number. See also "Simulation mode".
Prerequisite "Simulation Mode" = Alarm/Warning
Value at switch-on 484 (Simulation active)
14.18 Diagnosis → Reset
Enter reset code (124)
Navigation Diagnosis → Reset → Reset (124)
Write permission Operators/Service engineers/Expert
Description Reset parameters completely or partially to the factory values or order configuration by
entering a reset code, see "Resetting to factory settings (reset)" section → 38.