Commissioning Deltabar FMD71, FMD72
46 Endress+Hauser
9.5.2 Configuring language via operating tool (FieldCare)
A0017654-EN
9.6 Measuring mode selection
L
WARNING
Changing the measuring mode affects the span (URV)
This situation can result in product overflow.
If the measuring mode is changed, the setting for the span (URV) must be checked in
the "Setup" operating menu and readjusted if necessary.
Measuring mode (005)
Navigation Setup → Measuring mode
Write permission Operator/Service engineers/Expert
Description Select the measuring mode.
The operating menu is structured differently depending on the measuring mode selected.
Options Pressure
Level
Factory setting Level
9.7 Selecting the high-pressure side
9.7.1 Defining the high-pressure side
High press. side (183)
Navigation Setup → High press. side

Table of Contents