Commissioning Deltabar FMD71, FMD72
64 Endress+Hauser
Navigation Expert → System → Management → Download select.
Write permission Operators/Service engineers/Expert
Description Selection of data packages for up/download function in Fieldcare and PDM.
Prerequisite DIP switch set to "SW" and "Damping" set to "on". If you download using the factory setting
"Configuration copy", all parameters required for a measurement will be downloaded. The
functionality of the "Electronics replace" setting is reserved for Endress+Hauser Service and
can be accessed only if the correct device access code is entered.
Options • Configuration copy: This option overwrites general configuration parameters with the
exception of the serial number, order number, calibration, pos. zero adjust, application
and day information.
• Device replacement: This option overwrites general configuration parameters with the
exception of the serial number, order number, calibration and position adjustment.
• Electronics replace: This option overwrites general configuration parameters.
Factory setting Copy configuration
9.14 Configuring the local display
9.14.1 Adjusting the local display
The local display can be adjusted in the following menu:
Display/operat. → 74
9.15 Protecting settings from unauthorized access
The settings can be protected from unauthorized access in two ways:
• Locking via write protection switch (hardware locking) → 30
• Locking via parameter (software locking) → 37