Deltabar FMD71, FMD72 Technical data
Endress+Hauser 119
15.3.5 Application limits
A high ratio between the level and head pressure or between the differential pressure and
static pressure can result in large measured errors. A maximum ratio of 1:10 is
recommended. For calculation purposes, please use the free "Applicator" calculation tool,
which is available online at "www.endress.com/applicator" or on CD-ROM.