Deltabar FMD71, FMD72 Description of Device Parameters
Endress+Hauser 89
Prerequisite "Level selection" = "In height"
"Calibration mode" = Dry -> entry
"Calibration mode" = Wet -> display
Factory setting 0.0
Full calib. (012/031)
Navigation Setup → Extended setup → Level → Full calib. (012/031)
Write permission Operators/Service engineers/Expert
Description Enter the output value for the upper calibration point (vessel full). The unit defined in "Unit
before lin." must be used.
Note In the case of wet calibration, the level (vessel full) must actually be available. The
associated pressure is then automatically recorded by the device.
In the case of dry calibration, the level (vessel full) does not have to be available. For the
"In pressure" level selection, the associated pressure must be entered in the "Full pressure"
parameter. The associated height has to be entered in the "Full height" parameter for the
"In height" level selection.
Factory setting 100.0
Full pressure (032)/(187)
Navigation Setup → Extended setup → Level → Full pressure (032)/(187)
Write permission Operators/Service engineers/Expert
Description Enter the pressure value for the upper calibration point (vessel full). See also "Full calib.".
Prerequisite "Level selection" = In pressure
"Calibration mode" = Dry -> entry
"Calibration mode" = Wet -> display
Factory setting URL of the sensor module
Full height (033)/(188)
Navigation Setup → Extended setup → Level → Full height (033)/(188)
Write permission Operators/Service engineers/Expert
Description Enter the height value for the upper calibration point (vessel full). The unit is selected via
the "Height unit" parameter.