Proline Promag P 500 HART Commissioning
Endress+Hauser 97
Corrected volume flow unit
→ 97
Corrected volume unit
→ 97
Parameter overview with brief description
Parameter Prerequisite Description Selection Factory setting
Volume flow unit – Select volume flow unit.
Result
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Unit choose list Country-specific:
• l/h
• gal/min (us)
Volume unit – Select volume unit. Unit choose list Country-specific:
• m³
• gal (us)
Conductivity unit The On option is selected in
the Conductivity
measurement parameter
parameter.
Select conductivity unit.
Effect
The selected unit applies for:
• Current output
• Frequency output
• Switch output
• Simulation process variable
Unit choose list µS/cm
Temperature unit – Select temperature unit.
Result
The selected unit applies for:
• Temperature parameter
• Maximum value parameter
• Minimum value parameter
• External temperature
parameter
• Maximum value parameter
• Minimum value parameter
Unit choose list Country-specific:
• °C
• °F
Mass flow unit – Select mass flow unit.
Result
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Unit choose list Country-specific:
• kg/h
• lb/min
Mass unit – Select mass unit. Unit choose list Country-specific:
• kg
• lb
Density unit – Select density unit.
Result
The selected unit applies for:
• Output
• Simulation process variable
Unit choose list Country-specific:
• kg/l
• lb/ft³
Corrected volume flow unit – Select corrected volume flow
unit.
Result
The selected unit applies for:
Corrected volume flow
parameter (→ 137)
Unit choose list Country-specific:
• Nl/h
• Sft³/h
Corrected volume unit – Select corrected volume unit. Unit choose list Country-specific:
• Nm³
• Sft³