EasyManua.ls Logo

Endress+Hauser Levelflex FMP54 - Page 177

Endress+Hauser Levelflex FMP54
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus Operating menu
Endress+Hauser 177
"Level" submenu
The Level submenu (→  177) is only visible for Operating mode (→  159) =
Level
Navigation  Setup → Advanced setup → Level
Medium type
Navigation  Setup → Advanced setup → Level → Medium type
Description Specify type of medium.
User interface Liquid
Solid
Factory setting FMP50, FMP51, FMP52, FMP53, FMP54, FMP55: Liquid
Additional information The Solid option is only available for Operating mode (→  159) = Level
This parameter determines the value of several other parameters and strongly
influences the complete signal evaluation. Therefore, it is strongly recommended not
to change the factory setting.
Medium property
Navigation  Setup → Advanced setup → Level → Medium property
Prerequisite Operating mode (→  159) = Level
EOP level evaluationFix DC
Description Specify relative dielectric constant ε
r
of the medium.
Selection Unknown
DC 1.4 ... 1.6
DC 1.6 ... 1.9
DC 1.9 ... 2.5
DC 2.5 ... 4
DC 4 ... 7
DC 7 ... 15
DC > 15
Factory setting Dependent on Medium type (→  177) and Medium group (→  160).