5.4.2.2 Last active diagnostic message
SENSOR_X_LAST_DIAG_NE107_STATUS
SENSOR_X _LAST_DIAG_TIMESTAMP
5.5 Communication related diagnostics
5.5.1 Bus termination
Each Modbus segment must be terminated at each end using a bus terminator resistor.
The device has a terminating resistor with on/off switch. The status of this terminating resistor is displayed
via the T-LED on the communication module and can also be read via the device menu or Modbus.
Path: Menu/Expert/Diagnostics/System information/Modbus
0 = RS485 terminating resistor is OFF
1 = RS485 terminating resistor is ON
5.5.2 Protocol counters
The device monitors communication and makes error counters available. The error counters can be read
via Modbus (8.3.3). The error counters may also be displayed directly at the device:
Path: Menu/Expert/Diagnostics/System information/Modbus