SD01189C/07/EN/04.17
Endress+Hauser
57
Register
Data Type
Bytes
R/W
Unit
Description
Selection:
0: Not available (default)
1: Available
8580
UNSIGNED8
1
R
None
Cabinet heating
Selection:
0: Available
1: Not available (default)
8581
UNSIGNED8
1
R
None
Hose heating filter
Selection:
0: Available
1: Not available (default)
8582
UNSIGNED8
1
R
None
Hose heating analyzer
Selection:
0: Available
1: Not available (default)
8583
UNSIGNED8
1
R
None
Ambient temperature sensor
Selection:
0: Available
1: Not available (default)
8.5.2 Analyzer SP 1 setup (Holding, FC: 03, 16, 23)
Register
Data Type
Bytes
R/W
Unit
Description
8500
UNSIGNED8
1
R/W
None
Start condition of first sample preparation cleaning
Selection:
0: Immediate (default)
1: Date/time
2: Disabled
8501
DATETIME
8
R/W
None
Date of first sample preparation cleanig
Factory default: 1970-01-01T00:00:00
8505
DATETIME
8
R/W
None
Time of first sample preparation cleanig
Factory default: 1970-01-01T00:00:00
8509
UNSIGNED32
4
R/W
None
Interval for sample preparation cleaning
Factory default: 1970-01-02T00:00:00
8511
UNSIGNED32
4
R/W
None
Duration of sample preparation cleaning
Factory default: 1970-01-01T00:01:00
8513
UNSIGNED32
4
R/W
None
Max. transport time of sample preparation
Factory default: 1970-01-01T00:15:00
8515
FLOAT
4
R/W
Celsius
Switch on temperature for sample preparation hose heating
Factory default: 5.0
8517
FLOAT
4
R/W
Celsius
Switch on temperature for outlet sample preparation cabinet heating
Factory default: 5.0
8519
FLOAT
4
R/W
Celsius
Switch on temperature for inlet sample preparation cabinet heating
Factory default: 10.0
8521
UNSIGNED32
4
R/W
None
Warn level for sample prepartion filter change
Factory default: 1970-03-02T00:00:00
8523
UNSIGNED32
4
R/W
None
Warn level for sample prepartion hose change
Factory default: 1970-03-02T00:00:00
8525
UNSIGNED8
1
R/W
None
Number of consecutive measurements for one channel
Factory default: 1
8526
UNSIGNED8
1
R/W
None
Limit change filter
Selection:
0: Off (default)
1: On
8527
UNSIGNED8
1
R/W
None
Limit change hose
Selection:
0: Off
1: On (default)