SD01189C/07/EN/04.17
Endress+Hauser
45
Current state of the cleaning program.
Selection:
0: Off (default)
1: Waiting
2: Waiting
3: Pre-rinse
4: Cleaning
5: Postrinsing
6: Failsafe
7: Manual mode
8: Waiting for execution
8.3.4 Configuration (Holding, FC: 03, 16, 23)
Temperature unit (0 = °C; 1 = °F, 2 = K)
Modbus setting : baud rate
Selection:
0: 1200
1: 2400
2: 4800
3: 9600
4: 19200 (default)
5: 38400
6: 57600
7: 115200
Modbus setting : address mode (1=Hardware, 0=Software)
Factory default: false
Modbus setting : parity
Selection:
0: Even (1 Stopbit) (default)
1: Odd (1 Stopbit)
2: None (2 Stopbit)
Activation of modbus interface
Selection:
0: Off
1: On (default)
Modbus setting : operating mode
Selection:
0: RTU (default)
1: ASCII
Timeout for mobdus communication watchdog
Factory default: 5.0
Command for controlling modbus statistics
Selection:
0: No operation (default)
1: Retransmitt error counters
2: Reset error counters
Activation of ethernet interface
Selection:
0: Off
1: On (default)
Ethernet setting: IP address of the device
Ethernet setting: Netmask
Factory default: 4294967040
Ethernet setting: IP address of the gateway
Ethernet statistics: Link settings