EasyManua.ls Logo

Endress+Hauser Liquiline CM44 series - Analyzer Program (Holding, FC: 03, 16, 23)

Endress+Hauser Liquiline CM44 series
75 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SD01189C/07/EN/04.17
Endress+Hauser
63
Register
Data Type
Bytes
R/W
Unit
Description
9297
UNSIGNED8
1
R
None
Turbid sample mode
Selection:
0: Off
1: On (default)
2: Automatic
9298
FLOAT
4
R
KiloGramPerCubicMeter
Concentration of the calibration standard
9300
UNSIGNED8
1
R
None
Start condition for calibration
Selection:
0: Immediate
1: Date/time
2: Disabled (default)
9301
UNSIGNED32
4
R
None
Interval for calibration
Factory default: 1970-01-03T00:00:00
9303
UNSIGNED8
1
R
None
Start condition for cleaning
Selection:
0: Immediate
1: Date/time
2: Disabled (default)
9304
UNSIGNED32
4
R
None
Interval for cleaning
Factory default: 1970-01-03T00:00:00
8.5.7 Analyzer Program (Holding, FC: 03, 16, 23)
Register
Data Type
Bytes
R/W
Unit
Description
9200
UNSIGNED8
1
R/W
None
Start condition for measurment
Selection:
0: Immediate (default)
1: Date/time
2: Continuous
9201
DATETIME
8
R/W
None
Start of first measurement : date
Factory default: 1970-01-01T00:00:00
9205
DATETIME
8
R/W
None
Start of first measurement : time
Factory default: 1970-01-01T00:00:00
9209
UNSIGNED32
4
R/W
None
Interval for measurement
Factory default: 1970-01-01T00:15:00
9211
UNSIGNED8
1
R/W
None
Start condition for calibration
Selection:
0: Immediate
1: Date/time
2: Disabled (default)
9212
DATETIME
8
R/W
None
Start of first calibration : date
Factory default: 1970-01-01T00:00:00
9216
DATETIME
8
R/W
None
Start of first calibration : time
Factory default: 1970-01-01T00:00:00
9220
UNSIGNED32
4
R/W
None
Interval for calibration
Factory default: 1970-01-03T00:00:00
9222
UNSIGNED8
1
R/W
None
Start condition for cleaning
Selection:
0: Immediate
1: Date/time
2: Disabled (default)
9223
DATETIME
8
R/W
None
Start of first cleaning : date
Factory default: 1970-01-01T00:00:00
9227
DATETIME
8
R/W
None
Start of first cleaning : time
Factory default: 1970-01-01T00:00:00
9231
UNSIGNED32
4
R/W
None
Interval for cleaning
Factory default: 1970-01-03T00:00:00