SD01189C/07/EN/04.17
Endress+Hauser
62 © Copyright Dec. 2016
Register
Data Type
Bytes
R/W
Unit
Description
29: Custom
30: Rinsing
31: Rinsing
32: Preparing
9214
UNSIGNED32
4
R
None
Remaining time of current activity
9216
FLOAT
4
R
Seconds
Operating time of filter mat
9218
FLOAT
4
R
Seconds
Operating time of peltier element
9220
UNSIGNED8
1
R
None
Type of sample preparation
Selection:
0: None (default)
1: Conductive
2: Capacitive
9221
UNSIGNED8
1
R
None
Availability of sample preparation
Selection:
0: SP1 (default)
1: SP2
2: SP1 + SP2
9222
UNSIGNED8
1
R
None
SP InstallationType
Selection:
0: Others (default)
1: Pressurized pipe sampling
9223
OCTETSTRING32
32
R
None
Measuring parameter
9239
UNSIGNED32
4
R
None
Number of calibrations
9241
DATETIME
8
R
None
Time stamp of last calibration
Factory default: 1970-01-01T00:00:00
9245
UNSIGNED8
1
R
None
Calibration method of last calibration
Selection:
0: None (default)
1: Zero point
2: Cal. factor
3: 2-point calibration
4: Application calibration
5: Numeric input zero point
6: Numeric input cal. factor
7: Numeric input appl. cal.
9246
FLOAT
4
R
None
Raw value
9248
UNSIGNED8
1
R
None
Raw value state
9249
FLOAT
4
R
KiloGramPerCubicMeter
Zero point
9251
UNSIGNED32
4
R
None
Date of last zero point calibration
9253
FLOAT
4
R
None
Calibration factor of the photometer
Factory default: 1.00
9255
UNSIGNED32
4
R
None
Date of last calibration factor determination
9257
UNSIGNED8
1
R
None
Application calibration active
9258
UNSIGNED8
1
R
None
Internal dilution factor
Factory default: 1
9259
FLOAT
4
R
None
External dilution factor
Factory default: 1.0
9261
FLOAT
4
R
KiloGramPerCubicMeter
Offset
9263
FLOAT
4
R
KiloGramPerCubicMeter
SP1: Measured value
9265
UNSIGNED8
1
R
KiloGramPerCubicMeter
SP1: Measured value state
9266
UNSIGNED32
4
R
None
SP1: Date of last measured value
Factory default: 1970-01-01T00:00:00
9268
UNSIGNED32
4
R
None
SP1: Time of last measured value
Factory default: 1970-01-01T00:00:00
9270
FLOAT
4
R
KiloGramPerCubicMeter
SP2: Measured value
9272
UNSIGNED8
1
R
KiloGramPerCubicMeter
SP2: Measured value state
9273
UNSIGNED32
4
R
None
SP2: Date of last measured value
Factory default: 1970-01-01T00:00:00
9275
UNSIGNED32
4
R
None
SP2: Time of last measured value
Factory default: 1970-01-01T00:00:00
9277
OCTETSTRING32
32
R
None
Measuring range
9293
UNSIGNED32
4
R
None
Digestion time
9295
FLOAT
4
R
Celsius
Digestion temperature