SD01189C/07/EN/04.17
Endress+Hauser
56 © Copyright Dec. 2016
8.5 Analyzer
8.5.1 Analyzer SP 1 Info (Input, FC: 04)
Ordercode of sample preparation
Extended ordercode of sample preparation
Serial number of sample preparation
Hardware version of sample preparation
Software version of sample preparation
Type of sample preparation
Operating mode
Selection:
0: Local control
1: Remote control (default)
Current state of sample preparation
Selection:
0: Inactive (default)
1: Sampling
2: Cleaning
3: Flush with air
4: Flush with cleaner
5: Flush filter
6: Flush with air
7: Flush with cleaner
8: Flush with air
9: Flush with cleaner
10: Cleaning up
11: Cleaning up
12: Cleaning up
13: Cleaning up
14: Flush with air
15: Flush with cleaner
16: Flush with air
17: Flush with cleaner
18: Flush filter
19: Sampling
20: Pump backward
21: Cleaning up
Operating hours of sample preparation filter
Factory default: 1970-01-01T00:00:00
Operating hours of sample preparation hose
Factory default: 1970-01-01T00:00:00
Operating hours of sample preparation membrane pump
Factory default: 1970-01-01T00:00:00
Operating mode
Selection:
0: Independent
1: Controlled (default)
Type of sample preparation
Selection:
0: 3rd party (default)
1: Inlet
2: Aeration/Outlet
3: Unknown
Pressurized air available
Selection:
0: Available (default)
1: Not available