EasyManua.ls Logo

Endress+Hauser Liquiline CM44 series - Analyzer Manual Actions (Holding, FC: 03, 16, 23); Analyzer Photometer Info (Input, FC: 04); Analyzer Measurement Data (Input, FC: 04); Analyzer Manual Actions State (Input, FC: 04)

Endress+Hauser Liquiline CM44 series
75 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SD01189C/07/EN/04.17
Endress+Hauser
64 © Copyright Dec. 2016
8.5.8 Analyzer Manual actions (Holding, FC: 03, 16, 23)
Register
Data Type
Bytes
R/W
Unit
Description
9400
UNSIGNED8
1
R/W
None
Mode change request
Selection:
0: None (default)
1: Enter fieldbus mode
2: Continue automatic mode
3: Restart automatic mode
9401
UNSIGNED8
1
R/W
None
Abort current activity
Selection:
0: None (default)
1: Abort current action
9402
UNSIGNED8
1
R/W
None
Manual activity request
Selection:
0: None (default)
1: Start measurement
2: Start calibration
3: Start cleaning
8.5.9 Analyzer Photometer Info (Input, FC: 04)
Register
Data Type
Bytes
R/W
Unit
Description
9600
OCTETSTRING16
16
R
None
Hardware version of photometer
9608
OCTETSTRING16
16
R
None
Software version of photometer
9616
FLOAT
4
R
Seconds
Operating hours of photometer
9618
FLOAT
4
R
Celsius
Temperature of photometer cuvette
Factory default: 25
9620
FLOAT
4
R
None
Raw value of photometer
8.5.10 Analyzer Measurement data (Input, FC: 04)
Register
Data Type
Bytes
R/W
Unit
Description
9700
UNSIGNED8
1
R
None
Amount of available/selectable measurands
9701
OCTETSTRING16
16
R
None
Descriptor of available/selectable measurand
9709
OCTETSTRING16
16
R
None
Descriptor of available/selectable measurand
9717
OCTETSTRING16
16
R
None
Descriptor of available/selectable measurand
9725
OCTETSTRING16
16
R
None
Descriptor of available/selectable measurand
9733
OCTETSTRING16
16
R
None
Descriptor of available/selectable measurand
9741
OCTETSTRING16
16
R
None
Descriptor of available/selectable measurand
9749
OCTETSTRING16
16
R
None
Descriptor of available/selectable measurand
9757
OCTETSTRING16
16
R
None
Descriptor of available/selectable measurand
9765
OCTETSTRING32
32
R
None
Measuring range
9781
UNSIGNED8
1
R
None
Automatic range switch
Selection:
0: Manual (default)
1: Automatic
8.5.11 Analyzer manual actions state (Input, FC: 04)
Register
Data Type
Bytes
R/W
Unit
Description
9900
UNSIGNED8
1
R
None
State of mode change request interface (0=ready)
9901
UNSIGNED8
1
R
None
Manual action request state. 0 = ready.

Table of Contents

Other manuals for Endress+Hauser Liquiline CM44 series

Related product manuals