3.3.2 RTU
The data are transmitted in binary format and are therefore more compact.
0 = Broadcast
247 = Unicast
A delimiter of at least 3.5 characters must be maintained between two Modbus telegrams.
Copied from the
request into the
reply
3.3.3 TCP
As with RTU, the data are transmitted in binary format. Address and checksum used in RTU mode are
dispensed with. The Modbus ADU is embedded in the TCP/IP telegram.
Number of
bytes
to follow
Identifies the transaction
Copied from the
request into the
reply
Copied from the
request into the
reply
Number of bytes to
follow
Number of bytes to follow
Identifies a slave that is
connected to a serial bus
located downstream.
Note
The devices in the Liquiline
product family do not use the
unit identifier. This should
always be set to 0.
Copied from the
request into the
reply
Via the TCP port 502, you can use up to 3 parallel TCP connections with the device simultaneously.
Please note the correct structure of the MBAP header.