SD01189C/07/EN/04.17
Endress+Hauser
58 © Copyright Dec. 2016
Timeout no sample
Factory default: 1970-01-01T00:05:00
Timeout no sample (pressurized pipe)
Factory default: 1970-01-01T00:01:00
Filter air cleaning duration
Factory default: 30.0
Filter air cleaning interval
Factory default: 1970-01-01T02:00:00
Filter air cleaning mode
Selection:
0: On (default)
1: Off
Cleaner residence time
Factory default: 1970-01-01T00:01:00
Cleaning discard time
Factory default: 1970-01-01T00:01:30
Pressurized pipe cleaning duration
Factory default: 10.0
Pressurized pipe cleaning discard time
Factory default: 180.0
Start condition of first pressurized pipe cleaning
Selection:
0: On (default)
1: Date/time
2: Off
Interval for pressurized pipe cleaning
Factory default: 1970-01-01T00:30:00
8.5.3 Analyzer SP 2 Info (Input, FC: 04)
Ordercode of sample preparation
Extended ordercode of sample preparation
Serial number of sample preparation
Hardware version of sample preparation
Software version of sample preparation
Type of sample preparation
Operating mode
Selection:
0: Local control
1: Remote control (default)
Current state of sample preparation
Selection:
0: Inactive (default)
1: Sampling
2: Cleaning
3: Flush with air
4: Flush with cleaner
5: Flush filter
6: Flush with air
7: Flush with cleaner
8: Flush with air
9: Flush with cleaner
10: Cleaning up
11: Cleaning up
12: Cleaning up
13: Cleaning up
14: Flush with air
15: Flush with cleaner
16: Flush with air
17: Flush with cleaner
18: Flush filter