EasyManua.ls Logo

Endress+Hauser Liquiline CM44 series - Sensor Conductivity (Input, FC: 04)

Endress+Hauser Liquiline CM44 series
75 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SD01189C/07/EN/04.17
Endress+Hauser
66 © Copyright Dec. 2016
Offset
Data Type
Bytes
R/W
Unit
Description
6: Cond c
7: Cond i
8: ORP
9: TU/TS
10: Nitrate
11: ISE
12: SAC
13: Ultrasonic interface
14: Chlorine
18: pH/ORP
19: Cond c 4-pol
20: Oxygen (opt. Memosens coupling)
21: TU
22: Inline photometer
8.6.2 Sensor Conductivity (Input, FC: 04)
Register
Data Type
Bytes
R/W
Unit
Description
0
DATETIME
8
R
None
Current sensor-diagnosis : time of diagnosis apperance
4
DATETIME
8
R
None
Manufacturing date of the sensor
8
DATETIME
8
R
None
Initial-operation date of sensor
12
DATETIME
8
R
None
Date of the last successful slope calibration of the sensor
16
FLOAT
4
R
None
Cell constant
18
FLOAT
4
R
None
Difference of cell constant between current and last calibration
20
FLOAT
4
R
DeltaCelsius
Temperature offset
22
FLOAT
4
R
None
Min. measured value
24
FLOAT
4
R
Celsius
Min. measured temperature value
26
FLOAT
4
R
None
Max. measured value
28
FLOAT
4
R
Celsius
Max. measured temperature value
30
FLOAT
4
R
Seconds
Operating hours of connected sensor
32
FLOAT
4
R
Seconds
Raw value
34
FLOAT
4
R
Seconds
Operating time >80°C
36
FLOAT
4
R
Seconds
Operating time >100°C (Cond 4pol)
38
FLOAT
4
R
Seconds
Operating time >120°C (cond.)
40
FLOAT
4
R
Seconds
Operating time >125°C (ind.)
42
FLOAT
4
R
Seconds
Operating time >140°C (Cond 2pol)
44
FLOAT
4
R
Seconds
Operating time >150°C (Cond i, Cond 4pol)
46
FLOAT
4
R
Seconds
Operating time <100nS (cond.)
48
UNSIGNED16
2
R
None
Number of sterilisation cycles
49
UNSIGNED16
2
R
None
Number of clean in place cycles
50
UNSIGNED16
2
R
None
Current sensor-diagnosis : diagnosis code
51
UNSIGNED8
1
R
None
Current sensor-diagnosis : status according to NAMUR NE107
52
UNSIGNED8
1
R
None
Selected sensor type. The selected type has to match with the connected
sensor type.
Selection:
0: None (default)
1: Oxygen (amp.)
2: Oxygen (opt. fixed cable)
3: pH Glass
4: pH/ORP
5: pH ISFET
6: Cond c
7: Cond i
8: ORP
9: TU/TS
10: Nitrate
11: ISE
12: SAC
13: Ultrasonic interface
14: Chlorine
18: pH/ORP
19: Cond c 4-pol
20: Oxygen (opt. Memosens coupling)
21: TU
22: Inline photometer