Powerfail reaction
Selection:
0: Last mode (default)
1: Manual mode
Number of retries in case of failed calibration
Factory default: 2
Calibration factor of the photometer
Factory default: 1.00
Max. allowed deviation of calibration factor for
calibration
Factory default: 10
Min. allowed calibration factor
Factory default: 0.5
Max. allowed calibration factor
Factory default: 2.0
Concentration of the calibration standard
Concentration offset for concentration value calculation
Dilution factor for concentration value calculation
Factory default: 1.0
Max. allowed deviation of zero point for calibration
Min. allowed calibration zero point
Max. allowed calibration zero point
8.5.6 Analyzer Info (Input, FC: 04)
Extended ordercode of the device
Current extended ordercode of the device
Hardware version of the device
Software version of the device
Operation mode
Selection:
0: Manual (default)
1: Automatic
2: Fieldbus
Current activity
Selection:
0: None (default)
1: Measurement, sample preparation
2: Calibration
3: Cleaning
6: Emptying sample collector
9: Empty hoses
10: Initializing
11: Cleaning up
12: Cleaning sample preparation
13: Dispenser replacement to top
14: Dispenser replacement to bottom
15: Testing liquidmanager actuator
16: Bottle insertion
17: Bottle removal
18: Open valves
19: Close valves
20: Rinse reagents
21: Rinse with water
22: Commissioning
23: Empty reactor
24: Empty dosing unit
25: Empty dilution module
26: Rinse analyzer
27: Reactor heating test
28: Reactor pressure test