SD01189C/07/EN/04.17
Endress+Hauser
27
#
Current state
Condition Action
Next state
1
Manual
MODE_CHANGE_STATE:=3
ACTION_REQUEST_STATE:=2
Manual
2
Manual
Mode Automatic
MODE_CHANGE_STATE:=1
ACTION_REQUEST_STATE:=2
Automatic
3
Automatic
Mode Manual
MODE_CHANGE_STATE:=3
ACTION_REQUEST_STATE:=2
Manual
4
Automatic
MODE_CHANGE_REQUEST:=0
MODE_CHANGE_STATUS:=1,
ACTION_REQUEST_STATE:=2
Automatic
5
Automatic
MODE_CHANGE_REQUEST:=1
and mode change is not
possible
MODE_CHANGE_STATE:=3,
ACTION_REQUEST_STATE:=2
Automatic
6
Automatic
MODE_CHANGE_REQUEST:=1
and mode change is accepted
MODE_CHANGE_STATUS:=2,
ACTION_REQUEST_STATE:=2
Modechange -FB
7
Modechange-FB
Mode change finished
ACTION_REQUEST_STATE:=1,
ACTION_REQUEST_STATE:=0
Feldbus
8
Feldbus
See “Fieldbus status machine”
Feldbus
9
Feldbus
MODE_CHANGE_REQUEST:=2
MODE_CHANGE_STATUS:=2
Modechange -A
10
Feldbus
MODE_CHANGE_REQUEST:=3
MODE_CHANGE_STATE:=1,
ACTION_REQUEST_STATE:=2,
Time-controlled programs are
restarted with the current
time.
Modechange -A
11
Modechange -A
Mode changed finished
MODE_CHANGE_STATE:=1
ACTION_REQUEST_STATE:=2
Automatic