Register
Data Type
Bytes
R/W
Unit
Description
20: Oxygen (opt. Memosens coupling)
21: TU
22: Inline photometer
X003
OCTETSTRING32
32
R
None
Sensor description/TAG
X019
OCTETSTRING16
16
R
None
Serial number of the sensor
X027
OCTETSTRING16
16
R
None
Hardware version of connected sensor
X035
OCTETSTRING16
16
R
None
Software version of connected sensor
X043
OCTETSTRING32
32
R
None
Manufacturer of the sensor
X059
OCTETSTRING32
32
R
None
Ordercode of the sensor
X075
DATETIME
8
R
None
Manufacturing date of the sensor
X079
DATETIME
8
R
None
Initial-operation date of sensor
X083
FLOAT
4
R
None
Min. measured value
X085
FLOAT
4
R
Celsius
Min. measured temperature value
X087
FLOAT
4
R
None
Max. measured value
X089
FLOAT
4
R
Celsius
Max. measured temperature value
X091
UNSIGNED8
1
R
None
Current sensor-diagnosis : component
Selection:
0: General (default)
1: Temperature Input
2: Binary Input
3: Binary Output
4: Current Input
5: Current Output
6: Relay
7: Memosens Sensor
8: PID Controller
9: Cleaning
10: Limit switch
11: Mathematic Function
12: Binary Input Module
13: Binary Output Module
14: Sample preparation
X092
UNSIGNED8
1
R
None
Current sensor-diagnosis : name of module
Selection:
0: Slot 1 (default)
1: Slot 1
2: Slot 2
3: Slot 3
4: Slot 4
5: Slot 5
6: Slot 6
7: Slot 7
8: Sampler
9: Software
10: Analyzer
X093
UNSIGNED8
1
R
None
Current sensor-diagnosis : port
X094
UNSIGNED8
1
R
None
Current sensor-diagnosis : status according to NAMUR NE107
Bit 0: Failure (F)
Bit 1: Function check (C)
Bit 2: Maintenance (M)
Bit 3: Out of specification (S)
Bit 4-7: Reserved = 0
X095
UNSIGNED16
2
R
None
Current sensor-diagnosis : diagnosis code
X096
UNSIGNED32
4
R
None
Current sensor-diagnosis : time of diagnosis apperance
X098
UNSIGNED8
1
R
None
Number of additional active sensor diagnosis
X099
UNSIGNED8
1
R
None
Past sensor-diagnosis : component
Selection:
0: General (default)
1: Temperature Input
2: Binary Input
3: Binary Output
4: Current Input
5: Current Output
6: Relay

Table of Contents