EasyManua.ls Logo

Endress+Hauser Liquiline CM44 series - Page 73

Endress+Hauser Liquiline CM44 series
75 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SD01189C/07/EN/04.17
Endress+Hauser
73
Register
Data Type
Bytes
R/W
Unit
Description
29
UNSIGNED8
1
R
None
Calibration method
Selection:
0: None (default)
1: Numeric input
2: 1-point calibration
3: 2-point calibration
4: Multipoint Calibration
5: Table
7: Zero point
8: Slope air
9: Slope H2O
10: Slope variable
11: Zeropoint reference
12: Slope reference
13: Offset
14: Slope
15: with temp. comp
16: without temp. comp
17: Standard addition
18: Numeric input slope
19: Numeric input zero point
20: Numeric input 2-point
21: Factory calib.
22: Slope test gas
23: Value acquisition
30
UNSIGNED16
2
R
None
Number of calibrations
31
UNSIGNED16
2
R
None
Current sensor-diagnosis : diagnosis code
32
UNSIGNED8
1
R
None
Current sensor-diagnosis : status according to NAMUR NE107
33
UNSIGNED8
1
R
None
Selected sensor type. The selected type has to match with the connected
sensor type.
Selection:
0: None (default)
1: Oxygen (amp.)
2: Oxygen (opt. fixed cable)
3: pH Glass
4: pH/ORP
5: pH ISFET
6: Cond c
7: Cond i
8: ORP
9: TU/TS
10: Nitrate
11: ISE
12: SAC
13: Ultrasonic interface
14: Chlorine
18: pH/ORP
19: Cond c 4-pol
20: Oxygen (opt. Memosens coupling)
21: TU
22: Inline photometer
34
FLOAT
4
R
None
Raw value