Operation Liquistation CSF48
102 Endress+Hauser
Settings with a time-paced Basic program
Settings with the Basic program type with 1 bottle
Sampling mode = "Time paced CTCV"
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Sampling interval (for version
with vacuum pump or
peristaltic pump)
Sampling interval (for version
with sampling assembly)
00:01:00 to 99:59:00
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
00:00:10 to 99:59:00
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
Set the sampling interval.
Dosing volume (for version
with vacuum pump or
sampling assembly)
Sampling volume (for version
with peristaltic pump)
Vacuum pump:
20 to 350 ml
Peristaltic pump:
10 to 10000 ml
Sampling assembly:
10 to 1000 ml
Factory setting
Vacuum pump:
200 ml
Peristaltic pump:
100 ml
Sampling assembly:
200 ml
Set the dosing volume or the sampling volume.
In the version with a vacuum pump or sampling
assembly, the volume is taken from the setup and can
only be modified there.
The dosing accuracy and the repeatability of a sample
volume < 20 ml can vary, depending on the specific
application.
Multiplier
(for version with vacuum
pump or sampling assembly)
1 to 10
Factory setting
1
You can change the sampling volume using the
multiplier.
For example, if the dosing volume is set to 200 ml, set
the multiplier to 2 for a sampling volume of 400 ml. 2
samples are taken in succession when sampling.
Bottle change mode Options
Number of samples
•Time
•External signal
Factory setting
Number of samples
The bottle can be changed after a specific number of
samples, after a time or by an external signal.
If Bottle change mode Number of samples is selected:
Samples per bottle 1 to 9999
Factory setting
1
Set the number of samples. If the bottle is full
beforehand based on the calculated level, the system
prevents more samples being added to the bottle. Such
samples are logged as failed samples in the program
logbook. At the same time, the diagnostics message
"Overfill check" (F353) is also triggered.