Liquistation CSF48 Maintenance
Endress+Hauser 217
12 Maintenance
Process pressure and temperature, contamination, electrical voltage
Risk of serious or fatal injury
‣ De-energize the device before removing the rear panel.
‣ Power can be supplied to switching contacts from separate circuits. De-energize these
circuits before working on the terminals.
‣ If a sensor has to be removed during maintenance work, avoid hazards posed by pressure,
temperature and contamination.
Electrostatic discharge (ESD)
Risk of damaging the electronic components
‣ Take personal protective measures to avoid ESD, such as discharging beforehand at PE or
permanent grounding with a wrist strap.
‣ For your own safety, only use genuine spare parts. With genuine parts, the function,
accuracy and reliability are also ensured after maintenance work.
12.1 Recommended maintenance
Maintenance work has to be carried out at regular intervals to ensure the efficient operation of
the sampler.
The maintenance work comprises:
• Replacing the wear parts
• Cleaning the device
The cleaning intervals depend heavily on:
•The medium
• The ambient conditions of the sampler (dust etc.)
• The programming intervals
For this reason, adapt the cleaning intervals to your specific requirements but always ensure
that these cleaning tasks are performed regularly.
Replacing wear parts
Wear parts are replaced by Endress+Hauser Service at one- and two-year intervals. Please
contact your local sales center in this regard.
Endress+Hauser offers its customers a maintenance contract. With a maintenance
contract, you can increase your level of operational safety and relieve your operating staff
of some of their workload. Ask your Endress+Hauser Service Organization for detailed
information on maintenance contracts.