Technical data Liquistation CSF48
258 Endress+Hauser
15.10.6 Suction height
• Vacuum pump:
Max. 6 m (20 ft) or max. 8 m (26 ft), depending on the version
• Peristaltic pump:
Max. 8 m (26 ft)
15.10.7 Hose length
Max. 30 m (98 ft)
15.10.8 Sample supply to sampling assembly
• Minimum height difference: 0.5 m (1.6 ft)
• Maximum hose length: 5 m (16 ft)
• Material: EPDM black, 13 mm ID
15.10.9 Temperature control
Temperature sensors:
• Sampling compartment temperature
• Sample temperature (optional)
• Outside temperature (optional)
Cooling module:
• Sample temperature range: 2 to 20 ˚C (36 to 68 ˚F)
- Factory setting: 4 ˚C (39 ˚F)
• Automatic defrost system
• Cooling rate in accordance with Ö 5893 (Austrian standard):
4 liters of water at 20 ˚C cool down to 4 ˚C in less than 210 minutes
• Temperature constancy of sample at 4 ˚C over the operating temperature range of -15 to 40
˚C
(5 to 105 ˚F)
15.11 Environment
15.11.1 Ambient temperature range
15.11.2 Storage temperature
-20 to 60 ˚C (0 to 140 ˚F)
• With cooling module:
• Without cooling module:
• With ASA+PC or stainless steel
housing:
• With plastic polystyrene housing:
-20 to 40 ˚C (0 to 100 ˚F)
0 to 40 ˚C (32 to 100 ˚F)
-20 to 40 ˚C (0 to 100 ˚F)
0 to 40 ˚C (32 to 100 ˚F)