Electrical connection Liquistation CSF48
44 Endress+Hauser
6.4 Terminal assignment for input/output signals
The following signals can be configured for external sampler control:
Input signals
2 analog signals 0/4 to 20 mA
2 binary signals > 100 ms pulse width or edge
Signals of digital sensors with Memosens protocol (optional)
Output signals
2 Binary signals > 1 s pulse width or edge
The controller must be opened to allow the signal cable, sensor cable and optional relay to be
connected.
6.5 Connecting the sensors
6.5.1 Sensor types with Memosens protocol
a0016348
Fig. 51: Connection example for binary output with relay
Sensor types Sensor cable Sensors
Digital sensors without additional
internal power supply
CYK10 with plug connection
and inductive signal
transmission
•pH sensors
•ORP sensors
Combined sensors
Amperometric oxygen sensors
Conductive conductivity sensors
Chlorine sensors
Fixed cable Inductive conductivity sensors
Digital sensors with additional
internal power supply
Fixed cable Turbidity sensors
Sensors for interface measurement
Sensors for measuring the spectral absorption
coefficient (SAC)
Nitrate sensors
Optical oxygen sensors
Ion-sensitive sensors