Maintenance Liquistation CSF48
218 Endress+Hauser
12.2 Calibration
12.2.1 Sensors
Sensors with Memosens protocol are calibrated at the factory.
Users must decide whether the process conditions present require calibration during initial
commissioning.
Additional calibration is not required in many standard applications.
Calibrate the sensors at sensible intervals depending on the process.
All information on calibration is provided in BA01245C "Memosens".
12.2.2 Distribution arm
The position of the distribution arm is set at the factory.
It is only possible to calibrate the distribution arm in the version with multiple bottles.
The distribution arm must be calibrated if:
• The distribution arm motor has been replaced
• Error message "F236 Distribution arm" appears on the display
1. Select the number of bottles in the "Setup/Basic setup" menu.
2. Proceed as follows to calibrate the distribution arm:
3. Afterwards perform a distribution arm test in the "Diagnostics/System
test/Reset/Distribution arm" menu.
Path: Menu/Calibration active
Function Options Info
Distribution arm
Go to ref. point Action The reference run is started. The reference point is
in the middle at the front.
For versions with a distributor plate, the reference
point is at the arrow in the middle of the plate. For
versions with a distribution assembly the reference
point is between bottle number 1 and the last
bottle.
With Adjust you can correct the distribution arm if the unit does not move to the reference point correctly. Use
the two arrow keys to correct the position.