Operation Liquistation CSF48
154 Endress+Hauser
10.5 Outputs
10.5.1 Binary outputs
The basic version of the device always has two binary outputs.
Possible application
--> For outputting a manipulated variable to connected actuators
The binary output must be assigned in the program or subprogram before it can be
activated.
If Input mode Current is selected:
Meas. value format Factory setting
#.#
Specify the number of decimal places.
Damping 0 to 60 s
Factory setting
0 s
The damping causes a floating average curve of the
measured values over the time specified.
Path: Menu/Setup/Outputs
Function Options Info
OutputBinary
Function Options
•Off
•Event
• Limit value
• Diagnostics message
• Cleaning
(only for version with
sensors with the
Memosens protocol)
Factory setting
Off
The following functions depend on the option selected.
Function = "Off" switches off the function of the binary
output and means no further settings are required.
For Cleaning:
The outputs can only switch 100mA. A relay
must be added to control valves or motors.
If Function: Event is selected:
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Select the level change of the signal
Path: Menu/Setup/Inputs
Function Options Info