Operation Liquistation CSF48
154 Endress+Hauser
10.5 Outputs
10.5.1 Binary outputs
The basic version of the device always has two binary outputs.
Possible application
--> For outputting a manipulated variable to connected actuators
The binary output must be assigned in the program or subprogram before it can be
activated.
If Input mode Current is selected:
Meas. value format Factory setting
#.#
Specify the number of decimal places.
Damping 0 to 60 s
Factory setting
0 s
The damping causes a floating average curve of the
measured values over the time specified.
Path: Menu/Setup/Outputs
Function Options Info
OutputBinary
Function Options
•Off
•Event
Limit value
Diagnostics message
Cleaning
(only for version with
sensors with the
Memosens protocol)
Factory setting
Off
The following functions depend on the option selected.
Function = "Off" switches off the function of the binary
output and means no further settings are required.
For Cleaning:
The outputs can only switch 100mA. A relay
must be added to control valves or motors.
If Function: Event is selected:
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Select the level change of the signal
Path: Menu/Setup/Inputs
Function Options Info