Diagnostics and troubleshooting Liquistation CSF48
198 Endress+Hauser
11.7 Pending diagnostic messages
The Diagnostics menu contains all the information on the device status.
Furthermore, various service functions are available.
The following messages are directly displayed every time you enter the menu:
"Most important message"
Diagnostics message recorded with the highest criticality level
"Past message"
Diagnostic message whose cause is no longer present.
All the other functions in the Diagnostics menu are described in the following chapters.
Diagnostics messages associated with sampling are deleted under the following conditions:
Diagnostics messages caused by sampling are deleted automatically with the next successful
sampling.
Diagnostics messages caused by the level of medium in the bottle are deleted the next time
the bottle is changed.
If diagnostics message "M313 Liquidsensor" appears five times in succession when running
a program, the active program is aborted for reasons of safety. This behavior on the part
of the device cannot be modified by deactivating the diagnostics message under
Menu/Setup/General settings/Diagnostics.
11.8 Diagnostics list
All the current diagnostics messages are listed here.
A time stamp is available for each message. Furthermore, the system also displays the
configuration and description of the message as saved in "Menu/Setup/General
settings/Diagnostics/Device behavior".
For this purpose, select the appropriate message and press the navigator.