Liquistation CSF48 Installation
Endress+Hauser 27
5.3 Sampling with a flow assembly
The sample is taken directly from the flow assembly which is installed in the sampler stand.
The flow assembly is used when sampling in pressurized systems, e.g.:
Containers located at a higher level
Pressurized pipes
Pumping with external pumps
The flow rate should be 1000 to 1500 l/h.
The outlet of the flow assembly must be unpressurized (e.g. drain, open channel).
a0013127
Fig. 19: Flow assembly
1
2
3
Flow assembly inlet: ¾"
Sample connection
Flow assembly outlet: 1¼"