Liquistation CSF48 Technical data
Endress+Hauser 253
Minimum load (typical)
Min. 100 mA with 5 V DC
Min. 1 mA with 24 V DC
Min. 5 mA with 24 V AC
Min. 1 mA with 230 V AC
15.8 Protocol-specific data
15.8.1 HART
Relay coupled with binary output
Switching voltage Load (max.) Switching cycles (min.)
230 V AC, cos = 0.8 to 1 5 A 100,000
24 V DC, L/R = 0 to 1 ms 5 A 100,000
Extension module
Switching voltage Load (max.) Switching cycles (min.)
230 V AC, cos = 0.8 to 1 0.1 A 700,000
2 A 120,000
115 V AC, cos = 0.8 to 1 0.1 A 1,000,000
2 A 170,000
24 V DC, L/R = 0 to 1 ms 0.1 A 500,000
2 A 150,000
Manufacturer ID 11
h
Device type 119C
h
(CM44x), 119D
h
(CSFxx)
Device revision 001
h
HART version 7.2
Device description files (DD/DTM) www.endress.com
Device Integration Manager (DIM)
Device variables 16 user-definable and 16 predefined device variables, dynamic variables
PV, SV, TV, QV
Supported features PDM DD, AMS DD, DTM, Field Xpert DD