Liquistation CSF48 Diagnostics and troubleshooting
Endress+Hauser 185
11.2 Diagnostic information on the onsite display
Up-to-date diagnostic events are displayed along with their status category, diagnostic code and
short text. By clicking the navigator, you can call up additional information and tips on remedial
measures.
11.3 Diagnostic information via web browser
The same diagnostic information as for the onsite display is available via the web server.
11.4 Diagnostic information via fieldbus
Diagnostic events, status signals and more information are transmitted according to the
definitions and technical capabilities of the respective fieldbus systems
11.5 Adjusting diagnostic information
11.5.1 Classification of diagnostics messages
More detailed information on the current diagnostics messages displayed is provided in the
DIAG/Diagnostics list menu.
In accordance with Namur specification NE 107, the diagnostics messages are characterized by:
• Message number
• Error category (letter in front of the message number)
– F = (Failure). A malfunction has been detected.
The cause of the malfunction is to be found in the sampling point/measuring point. Any
controller connected should be set to manual mode.
– M = Maintenance required. Action should be taken as soon as possible.
The device still measures/takes a sample correctly. Immediate measures must not be taken.
However, proper maintenance efforts would prevent a possible malfunction in the future.
– C = (Function check) (No error).
Maintenance work is being performed on the device. Wait until the work has been
completed.
– S = Out of specification. The measuring point is being operated outside specifications.
Operation is still possible. However, you run the risk of increased wear, shorter operating
life or lower accuracy levels. The cause of the problem is to be found outside the measuring
point.
• Message text
If you contact the Service Department, please cite the message number only. Since you can
individually change the assignment of an error to an error category, the Service Department
cannot use this information.