Diagnostics and troubleshooting Liquistation CSF48
186 Endress+Hauser
11.5.2 Adjusting the device behavior
All the diagnostics messages are assigned to specific error categories at the factory. Since other
settings might be preferred depending on the application, error categories and the effect errors
have on the measuring point can be configured individually. Furthermore, every diagnostics
message can be disabled.
Example
The controller returns diagnostics message 531 "Logbook full". You want to change this message
so that an error is not indicated on the display for example.
1. Go to:
– Menu/Setup/General settings/Extended setup/Diagnostics/Device behavior
for device-specific diagnostics messages (as in this example)
– Menu/Setup/Inputs/<Sensor type>/Extended setup/Diagnostics settings/Diag.
behavior
for sensor-specific diagnostics messages.
2. Select the diagnostics message and press the navigator button.
3. Decide:
a. Should the message be deactivated?
b. Do you want to change the error category?
c. Should an error current be output?
d. Do you want to trigger a cleaning program?
4. Deactivate the message, for example (Diagnostics message to "Off").
Configuration options
The list of diagnostic messages displayed depends on the path selected. There are
device-specific messages, and messages that depend on what sensor is connected.
Path: ... /Extended setup/Diagnostics settings/Diag. behavior (optional)
Function Options Info
List of diagnostic messages Select the message to be changed. Only then can you
make the settings for this message.
Diag. code Read only
Diagnostic message Options
•On
•Off
Factory setting
Depends on the message
You can deactivate or reactivate a diagnostic message
here.
Deactivating means:
• No error message in the measuring mode
• No error current at the current output