Liquistation CSF48 Diagnostics and troubleshooting
Endress+Hauser 187
Error current Options
•On
•Off
Factory setting
Depends on the message
Decide whether an error current should be output at the
current output if the diagnostic message display is
activated.
If general device errors occur, the error current is
output at all the current outputs.
In the case of channel-specific errors, the error
current is only output at the particular current
output.
Status signal Options
• Maintenance (M)
• Out of specification (S)
•Function check (C)
• Failure (F)
Factory setting
Depends on the message
The messages are divided into different error categories
in accordance with NAMUR NE 107.
Decide whether you want to change a status signal
assignment for your application.
Diag. output Options
•None
• Alarm relay
• Relay 1 to n (depends on
the device version)
Factory setting
None
You can use this function to select a relay output and/or
binary output to which the diagnostic message should be
assigned.
For sensors with the Memosens protocol:
Before you can assign the message to an output, you
must first configure a relay output to "Diagnostics"
(Menu/Setup/Outputs, assign the "Diagnostics" function
and set the Operating mode to "as assigned").
Cleaning
program(optional)
Options
•None
• Cleaning 1
• Cleaning 2
• Cleaning 3
• Cleaning 4
Factory setting
None
Decide whether the diagnostic message should trigger a
cleaning program.
You can define the cleaning programs under:
Menu/Setup/Additional functions/Cleaning.
Detail information Read only Here you can find more information on the diagnostic
message and instructions on how to resolve the
problem.
Path: ... /Extended setup/Diagnostics settings/Diag. behavior (optional)
Function Options Info