Operation Liquistation CSF48
112 Endress+Hauser
Settings with a time/flow-paced Basic program (only for version with peristaltic pump)
Settings with the Basic program type with 1 bottle
Sampling mode = "Time/flow paced CTVV"
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Sampling volume input Options
• No flow input configured
• Binary input S:x
• Current input S:x
Factory setting
No flow input configured
Select the sampling volume input.
The binary input or the current input must be
configured for this function.
Only the inputs configured as a sampling volume
input are displayed.
Sampling interval 00:01:00 to 99:59:00
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
Set the sampling interval.
If Sampling volume input Binary input is selected:
Sampling volume / pulse 10 to 1000 ml
Factory setting
20 ml
Set what sampling volume should be grabbed per
pulse.
The dosing accuracy and the repeatability of a sample
volume < 20 ml can vary, depending on the specific
application.
If Sampling volume input Current input is selected:
Sampling volume 20mA 10 to 10000 ml
Factory setting
100 ml
Set what sampling volume should be grabbed at 20
mA.
The dosing accuracy and the repeatability of a sample
volume < 20 ml can vary, depending on the specific
application.
Flow calculation Options
• Current
•Average flow
Factory setting
Current
Current:
The current flow is converted to the sampling volume
at the time of sampling.
Average flow:
The system calculates the mean between the last and
the current sample and sets the sampling volume
accordingly.
Bottle change mode Options
• Number of samples
•Time
•External signal
Factory setting
Number of samples
The bottle can be changed after a specific number of
samples, after a time or by an external signal.