Liquistation CSF48 Operation
Endress+Hauser 115
Bottle change mode Options
•Number of samples
•Time
•External signal
Factory setting
Number of samples
The bottle can be changed either after a specific
number of samples, after a time or by an external
signal.
If Bottle change mode Number of samples is selected:
Samples per bottle 1 to 9999
Factory setting
1
Set the number of samples.
If Bottle change mode Time is selected:
Time interval 00-00:02 to 31-00:00
DD-HH:MM
Factory setting
00-01:00 DD-HH:MM
Set the time (in days, hours and minutes) after which
the system should change to the next bottle.
Multiple bottles 0 to 23
The configuration options
depend on the current number
of bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to separate
bottles.
If Bottle change mode External signal is selected:
Bottle change signal input Options
• No bottle change input
configured
• Binary input S:x
Factory setting
No bottle change input
configured
The bottle change input can be configured under
Inputs .
Multiple bottles 0 to 23
The configuration options
depend on the current number
of bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to separate
bottles.
Start condition Options
•Immediate
•Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable time.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info