Operation Liquistation CSF48
148 Endress+Hauser
Unit of totalized flow
Current totalized flow - - - The totalized flow values are displayed.
Reset totalizer Options
•Manual
•Automatic
•At program start
Factory setting
Manual
Manual:
Reset the counter manually.
Automatic:
The counter is reset automatically at intervals.
At program start:
The counter is reset at program start.
If counter reset Manual is selected:
Reset totalized flow Action The totalized flow currently calculated is set to zero
when the counter is reset.
If counter reset Automatic is selected:
Interval Options
• Daily
•Weekly
•Monthly
Factory setting
Daily
Daily:
If a daily interval is selected, set the Time in the menu
item that follows.
Weekly:
If a weekly interval is selected, set the Day of week and
the Time in the menu items that follow.
Monthly:
If a monthly interval is selected, set the Day of month
and the Time in the menu items that follow.
Time 00:00:00 to 23:59:59
HH:MM:SS
Factory setting
12:00:00 HH:MM:SS
If Input mode Rainfall is selected:
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Preselect the level change of the signal.
Unit Options
•mm
•inch
Factory setting
mm
Select the unit.
Meas. value format Factory setting
#.#
Specify the number of decimal places.
1 Impulse = 0.00 to 5.00 mm
Factory setting
1.0 mm
Definition of the pulse value, limits are calculated
depending on the unit. The correct switch value is
provided in the Operating Instructions of your rain
gage.
Path: Menu/Setup/Inputs
Function Options Info