Operation Liquistation CSF48
156 Endress+Hauser
Operating mode Options
as assigned
•Namur M
•Namur S
•Namur C
•Namur F
Factory setting
as assigned
as assigned:
If this option is selected, the diagnostics messages
which you have individually assigned to the binary
output are output via the binary output.
Namur M - F:
If you decided to use one of the Namur classes, all the
messages that are assigned to the individual class are
output via the binary output.
You can also change the Namur class assignment for
every diagnostics message. (Menu/Setup/General
settings/Diagnostics/Device behavior or
Menu/Setup/Inputs/../Diagnostics settings/Diag.
behavior)
Attributed diagnostic
messages
Read only
List of diagnostic messages
All the messages assigned to the relay output appear
on the display. You do not have the option of editing
the information here.
If Function: Cleaning is selected: (only for version with sensors with the Memosens protocol)
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Select the level change of the signal
Assignment Options
•None
Cleaning 1-4
Factory setting
None
Use this function to choose the cleaning instance
which should be started when the binary output is
active.
Path: Menu/Setup/Outputs